30 October 2007: Prepared for OCB data system by Nancy Copley (Biology Dept, WHOI) and Cyndy Chandler, BCO-DMO (MCG Dept, WHOI).
Contact: Jim Ledwell (WHOI)
The tracer injection sled for the EDDIES project lowered to the target potential density surface and then towed along that surface at 0.5 m/s while the tracer is injected. The sled is neutrally buoyant and is towed at the end of a 2-meter tether attached to the end of the CTD cable, which removes much of the ship motion. To follow lower frequency displacements of the target surface, the winch is controlled automatically with feedback from a SeaBird 911plus CTD system (Ledwell et al., 1998). The CTD has dual pumped C/T sensors mounted at the front of the sled to sample water not perturbed by the thermal wake of the sled. Also with the CTD was the same WET Labs fluorometer that had been used on Survey 1 on the Oceanus CTD/Rosette system. On the injection sled are also mounted injection orifices, pumps, batteries, fluid reservoirs, and control electronics for the injection.
The sampling sled is similar to the injection sled in that it carries the same SeaBird 9 CTD and the WET Labs fluorometer. It is mounted directly at the end of the CTD cable, with no tether, and with an array of integrating samplers above and below it (Ledwell et al. 1998). It is oriented into the flow with wedge-shaped panels on the aft end, and is also controlled to stay as near as possible to a target isopycnal surface.
Ledwell, J. R., A. J. Watson, and C. S. Law, Mixing of a tracer in the pycnocline, J. Geophys. Res., 103 (C10), pp. 21,499-21,529, 1998.
Data from the casts with the injection and sampling sleds are divided into downcast, flights along the target surface, and upcasts, indicated by "down", "flight" and "up" in the filenames, respectively.
There are 2 injection casts, 003 and 004, and 7 sampler casts, 001 through 007.
The sled sampling events as reported in the cruise event log are:
# cruise: OC415-2 EDDIES Tracer 1 # event date time cast lon lat ev_type comments --------------------------------------------------------------------------- 200507202354 20050720 2354 nd nd nd Inject Test 200507211601 20050721 1601 1 -67.123 30.560 Inject Start Inject 200507211604 20050721 1604 1 -67.123 30.559 Inject End Inject 200507220001 20050722 0001 2 -67.001 30.595 Inject Start Inject 200507220018 20050722 0018 2 -67.003 30.591 Inject End Inject 200507230018 20050723 0018 3 -67.241 30.450 Inject Start Inject 200507231020 20050723 1020 3 -67.184 30.293 Inject End Inject 200507231540 20050723 1540 4 -67.271 30.308 Inject Start Inject 200507231800 20050723 1800 4 -67.259 30.265 Inject End Inject 200507262333 20050726 2333 1 -66.515 28.169 Sampler Sampler Start 200507270045 20050727 0045 1 -66.509 28.154 Sampler Sampler End 200507280032 20050728 0032 2 -67.429 30.233 Sampler Sampler Start 200507281202 20050728 1202 2 -67.190 30.017 Sampler Sampler End 200507290048 20050729 0048 3 -67.717 30.090 Sampler Sampler Start 200507291202 20050729 1202 3 -67.461 30.001 Sampler Sampler End 200507300024 20050730 0024 4 -67.567 29.934 Sampler Sampler Start 200507301201 20050730 1201 4 -67.618 30.161 Sampler Sampler End 200507310034 20050731 0034 5 -67.618 30.003 Sampler Sampler Start 200507311202 20050731 1202 5 -67.539 30.233 Sampler Sampler End 200508010127 20050801 0127 6 -67.653 29.911 Sampler Sampler Start 200508011230 20050801 1230 6 -67.642 30.159 Sampler Sampler End 200508020130 20050802 0130 7 -67.049 29.890 Sampler Sampler Start 200508021234 20050802 1234 7 -67.695 30.158 Sampler Sampler End
The primary variables added to the database from the original INJ_00?_*.mat files are:
scan scan number
yearday time in decimal days, referenced to 0 at start of 2005
temp0 temperature, primary sensor pair (T90, deg C))
temp1 temperature, secondary sensor pair (T90, deg C))
cond0 conductivity, primary sensor pair
cond1 conductivity, secondary sensor pair
press pressure (dbar)
v5 fluorometer voltage
pump_1 Tracer pump 1 status: 1=ON; 0=OFF;
pump_2 Tracer pump 2 status: 1=ON; 0=OFF;
and the derived variables are:
sal0 Salinity, primary sensor pair
sal1 Salinity, secondary sensor pair
fl Chlorophyll from fluorometer voltage, v5, (mg/m3):
fl = 3.5334*v5 - 0.3250, with v5 in volts
The primary variables added to the database from the original SAM_00?_*.mat files are:
scan scan number
yearday time in decimal days, referenced to 0 at start of 2005
lat decimal degrees north
lon decimal degrees east (negative in our case)
temp0 temperature, primary sensor pair (T90, deg C))
temp1 temperature, secondary sensor pair (T90, deg C))
cond0 conductivity, primary sensor pair
cond1 conductivity, secondary sensor pair
press pressure (dbar)
v5 fluorometer voltage
sampl Sampler status: 1=ON; 0=OFF
and the derived variables are:
sal0 Salinity, primary sensor pair
sal1 Salinity, secondary sensor pair
fluor_Wetl Chlorophyll from fluorometer voltage, v5, (mg/m3):
fluor = 3.5334*v5 - 0.3250, with v5 in volts
Salinity Calibration
A few salinity samples were taken during the tows during both OC415-2.
The mean values should be subtracted from sal0, and potential temperature and potential density should be recalculated accordingly.
Cruise | Name | No. of samples | Mean S - Autosal | Std. Dev. |
OC415-2 | Tracer 1 | 12 | 0.0069 | 0.0023 |
PI notes pertaining to the Seabird 911plus CTD unit
SeaBird 911plus set up parameters
24 scans were averaged, so data were recorded at 1 hz.
A correction for the thermal mass effect was made with coefficients:
a = 0.027300, b = 0.100000
Calibration coefficients used on this run:
conductivity sensor 0 Serial No = 224
g = -4.235789670000e+000
h = 5.030437210000e-001
i = -2.164846790000e-003
j = 1.373217080000e-004
CTcor = 3.250000000000e-006
CPcor = 3.616348326169e-230
conductivity sensor 1 Serial No = 763
g = -4.219668290000e+000
h = 4.468958890000e-001
i = 8.929849640000e-005
j = 1.422904140000e-005
CTcor = 3.250000000000e-006
CPcor = 3.616348326169e-230
temperature sensor 0: (ITS090) Serial No. = 1080
f0 = 1.000000e+003
g = 4.795938520000e-003
h = 6.727434680000e-004
i = 2.635684590000e-005
j = 2.154517740000e-006
temperature sensor 1: (ITS-90) Serial No. = 1085
f0 = 1.000000e+003
g = 4.858597320000e-003
h = 6.721669580000e-004
i = 2.508981820000e-005
j = 1.909927960000e-006
pressure Serial No = 59933
C1 = -4.462697000000e+004
C2 = -2.850590000000e-001
C3 = 1.241750000000e-002
D1 = 3.985600000000e-002
D2 = 0.000000000000e+000
pressure:
T1 = 3.041551000000e+001
T2 = -4.168600000000e-004
T3 = 3.885970000000e-006
T4 = 3.330870000000e-009
T5 = 0.000000000000e+000
Voltage Assignments:
V0 = n/a
V1 = n/a
V2 = n/a
V3 = n/a
V4 = n/a
V5 = WetLabs Fluorometer
V6 = n/a
V7 = n/a